PROBLEM
POSSIBLE CAUSE
POSSIBLE SOLUTION
The air intake on the frame
and/or the cover is
obstructed, or the heat
exchanger is not sufficiently
ventilated.
Remove all possible obstructions from
the air intake and heat exchanger.
Check that the device is not in direct
contact with walls or surfaces (see
Section 2 Location and Installation).
Vacuum test has failed
The unit chamber was hot
when the test was initiated.
Ensure stand-by chamber warming
system is turned OFF. (See Section 7.4
Setting the Stand-by Mode). Attempt a
second Vacuum Test making sure the
chamber is cooled to room
temperature.
Water is pooling under the unit
There is a leak in the system.
Contact your SciCan dealer.
Bowie-Dick test has failed
Attempt a second Bowie-Dick test. If it
fails, contact your SciCan dealer.
No cycles are stored in the
unit’s memory
Logic board configuration
issue
Check unit serial number to see if it
was accurately updated after a logic
board service. If number consists of
zeros, call SciCan.