47
Using Process Enforced Documentation mode:
This mode activates the process enforced usage function along with the additional documentation function
that generates a report with information about the cycle and the load type.
When starting a cycle with Process Enforced Documentation mode turned ON, you will be prompted to
identify the general contents of the load to be processed by selecting from a list as well as whether a
biological indicator and chemical indicator are included.
At the end of the cycle, you will be prompted to report whether the indicators have passed and whether
the load is dry (as applicable).