74
The device screen is divided into four groups (see image below).
Section A is for monitoring all systems. The values displayed here are generated in real time.
Section B is for checking the heating systems. To verify the functioning of either the steam generator (TSG)
or the top and bottom band heaters (TBHT and TBHB), you can use the arrows to the right to increase the
temperature in that component to a specific target temperature and then watch the corresponding reading
in Section A to see if the component successfully reaches the target temperature.
Section C is for checking the individual functioning of valves, pumps and the heat exchanger fan. Press on a
button to activate that individual component or press the buttons in a specific sequence to simulate an
operating condition.
Section D locks and unlocks the door. There are 3 microswitches that control the door functioning. Door
closed (DCL), door latched (DLA) and door locked (DLK) are listed in Section A along with their ON/OFF values
to show that the microswitches are functioning correctly.