Repairs
FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH 51
Subject to change without notice
3.5.9.2 Fit the Ex-e cable gland
1 Unpack the new cable gland and check for transport damage. Do not use the cable
gland when any damage can be seen.
2 Guide the cable gland through the
Ex-d terminal compartment.
3 Fit the snap ring.
4 Make sure the snap ring engages
properly in the groove.
5 Connect the electrics as described in §3.4 “Electrical Installation” in the
FLOWSIC600-XT Operating Instructions.
6 Close the Ex-e terminal compartment again, p. 49, § 3.5.8.4.