Repairs
FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH 53
Subject to change without notice
3.5.10.2 Fit the new cable gland
1 Unpack the new cable gland and check for transport damage. Do not use the cable
gland when any damage can be seen.
2 Position the cable gland. Use a small
screwdriver as aid when necessary.
3 Fix the main board with the 3 long
screws.
4 Fit the snap ring.
5 Make sure the snap ring engages
properly in the groove.
6 Refit the Ex-d I/O block, p. 44, §3.5.6.3.
7 Refit the SPU circuit board, p. 40, §3.5.5.3.