ID/
Icons
Diagnostic Action Comment
162 Density Value simulated. Disable 'Simulation' before re‐
turning to normal operation
163 Medium temperature Value simulated. Disable 'Simulation' before re‐
turning to normal operation
166 Standard volume flow Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Only hydrocarbon
167 Totalizer 1 Value simulated. Disable 'Simulation' before re‐
turning to normal operation
168 Totalizer 2 Value simulated. Disable 'Simulation' before re‐
turning to normal operation
169 Totalizer 3 Value simulated. Disable 'Simulation' before re‐
turning to normal operation
170 Loop current Loop current simulated. Disable 'Simulation' be‐
fore returning to normal operation
Only HART devices
172 Transmitter FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
173 Sensor FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)
110 Operating Instructions, 08/2018, A5E37367739-AD