ID/
Icons
Diagnostic Action Comment
206 CH3 Pulse overflow. Pulse output insufficient pulse
separation. Reduce pulses per amount, or re‐
duce pulse width, or increase amount
If Operation mode is config‐
ured to Pulse output
207 CH4 Loop current in lower saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 4 parameter 'Lower range value'
If Operation mode is config‐
ured to Current output
208 CH4 Loop current in upper saturation. Check proc‐
ess conditions or align limit to normal operation.
Adjust channel 4 parameter 'Upper range value'
If Operation mode is config‐
ured to Current output
209 CH4 Cable break. Check channel 4 current output
cable connection
If Operation mode is config‐
ured to Current output
210 CH4 Frequency too low. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 4 parameter 'Lower range value'
If Operation mode is config‐
ured to Frequency output
211 CH4 Frequency too high. Check process conditions
or align limit to normal operation. Adjust chan‐
nel 4 parameter 'Upper range value'
If Operation mode is config‐
ured to Frequency output
212 CH4 Pulse overflow. Pulse output insufficient pulse
separation. Reduce pulses per amount, or re‐
duce pulse width, or increase amount
If Operation mode is config‐
ured to Pulse output
214 CH2 Channel simulated. Disable 'Simulation' before
returning to normal operation
215 CH3 Channel simulated. Disable 'Simulation' before
returning to normal operation
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)