4.4.9 Analog output
Termination Clamp
Signal Comment
X1.29 IOUT+ Analogue output +
X1.30 IOUT- Analogue output -
Table 4-7 Analog output connection
Figure 4-8
Analog output
4.4.10 Interface RS 485 for Mettler Toledo type WM or WMH
Digital sensors from the Mettler Toledo company of type WM or WMH can
be connected to Siwarex. The following interface parameters (data record 3
and 7) are to be set:
(DR3) Device selection: Digital load cell Mettler Toledo
(DR7) Baudrate: 38400
(DR7) Character parity: even
(DR7) Data bits: 7
(DR7) Stop bits: 1
4-22 SIWAREX FTC