5.2.15 DR3 - Minimum Weight for Weighing Range 1
The numeral step for weighing range 1 can be defined corresponding with standard
EN 45501 (0.0001 to 50).
5.2.16 DR 3 - Minimum Weight for Weighing Range 2
The cu
rrent weight value can be used for verifiable recording (with the numeral
step for measurement range 2) only above the minimum weight. The minimum
weight is defined by an adjustment or a calibration. The minimum weight depends
on the number and type of load cells used. The value can be set to 0 on non-
calibrating scales.
Note
If only one weighing range is defined (such as e.g. for the belt scale or loss-in-
weight scale), then this definition is meaningless.
5.2.17 DR 3 - Maximum Weight for Weighing Range 2
The cu
rrent weight can be used for calibrating purposes (with the numeral step for
measurement range 2) only under the maximum weight. The maximum weight is
defined during adjustment.
The maximum weight depends on the number and type of load cells used.
Note
If only one weighing range is defined then this entry is meaningless.
5.2.18 DR 3 - Minimum Weight for Weighing Range 2
The num
eral step for weighing range 2 can be defined corresponding with standard
EN 45501 (0.0001 to 50).
If only one weighing range is defined then this entry is meaningless.
5.2.19 DR 3 - Minimum Weight for Weighing Range 3
The cu
rrent weight value can be used for verifiable recording (with the numeral
step for measurement range 3) only above the minimum weight. The minimum
weight is defined by an adjustment or a calibration. The minimum weight depends
on the number and type of load cells used.
The value can be set to 0 on non-calibrating scales.
Note
If only one or two weighing ranges are defined then this entry is meaningless.
5.2.20 DR 3 - Maximum Weight for Weighing Range 3
The cu
rrent weight can be used for calibrating purposes (with the numeral step for
measurement range 3) only under the maximum weight. The maximum weight is
defined during adjustment.
The maximum weight depends on the number and type of load cells used.
Note
If only one or two weighing ranges are defined then this entry is meaningless.
5-42 SIWAREX FTC