WEIGHING FUNCTIONS AND PARAMETERS
5.2.21
DR 3 - Numeral Step for Weighing Range 3
The num
eral step for weighing range 1 can be defined corresponding with standard
EN 45501 (0.0001 to 50).
Note
If only one or two weighing ranges are defined then this entry is meaningless.
5.2.22 DR 3 - Standstill Time
Standstill mo
nitoring is used to identify when the scale is at a stable equilibrium.
Scale standstill is determined if the weight value changes by less than a defined
deviation range (standstill value) within a defined time (standstill time). Standstill is
used in static operation of the scale (Command: Zero setting, Taring, Output log).
5.2.23 DR 3 - Standstill Range
Standstill mo
nitoring is used to identify when the scale is at a stable equilibrium.
Scale standstill is determined if the weight value changes by less than a defined
deviation range (standstill value) within a defined time (standstill time). Standstill is
used in static operation of the scale (Command: Zero setting, Taring, Output log).
T
he figure below illustrates the functioning of the standstill monitoring feature.
FIGURE 5-5
Standstill Monitoring
5.2.24 DR 3 - Waiting Time for Standstill
To preve
nt having to cancel a weighing instruction (only taring, printing and zero
setting in static operation) immediately when the stable weight of the scale is not
defined, the waiting time is given as standstill. If the scale instruction tare, output
SIWAREX FTC