plausibility test corresponding with the defined operating mode. If there is a
parameter error then the data record will not be accepted (stored) by the
SIWAREX FTC and a "synchronous" message is generated (see chapter
7
Messages and Diagnosis
).
5.2 DR3 Adjustment Parameters
The adjustment parameters must be checked for each new scale and must be
changed where necessary and most important, the operating mode must be
defined during commissioning.
The scale is basically defined using the adjustment parameter and by performing
the adjustment. Adjusting the scale can be done using adjustment weights
(see
5.2.1) or theoretically (see 5.3). In verifiable operation as a non-automatic
weighing machine, the data of the DR3 cannot be changed after calibration.
Procedures:
o Check all parameters and change if necessary
o Switch on service operation (command).
o Send the DR3 to the scale
o Perform scale adjustment
o Receive DR3 from the scale
o Switch off service operation
Name Type Address Default Range of Values / Meaning Reference
(6) Adjustment
Adjustment digits for
Adjustment zero 0
DINT DBD000 1398101 0 to: 2
24
Other definitions not permitted.
5.2.1
Adjustment digits for
Adjustment weight 1
DINT DBD004 15379113 0 to: 2
24
Other definitions not permitted.
5.2.1
Adjustment digits for
adjustment weight 2
DINT DBD008 0 0 to: 2
24
Other definitions not permitted.
5.2.1
Adjustment digits for
Adjustment weight 3
DINT DBD012 0 0 to: 2
24
Other definitions not permitted.
5.2.1
Adjustment digits for
Adjustment weight 4
DINT DBD016 0 0 to: 2
24
Other definitions not permitted.
5.2.1
Adjustment weight 1
(Flow rate with bulk flow
measurement device)
REAL DBD020 50 > 0
Other definitions not permitted.
5.2.1
adjustment weight 2
(Flow rate with bulk flow
measurement device)
REAL
DBD024 0 0
Other definitions not permitted.
5.2.1
Adjustment weight 3
(Flow rate with bulk flow
measurement device)
REAL DBD028 0 0
Other definitions not permitted.
5.2.1
Adjustment weight 4
(Flow rate with bulk flow
measurement device)
REAL
DBD032 0 0
Other definitions not permitted.
5.2.1
5-32 SIWAREX FTC