HARDWARE CONFIGURATION AND ASSEMBLY
Certificate: c-UL-us haz. loc.
UL file number. USA: NRAG.E239877
UL file number Canada: NRAG7.E239877
Note to c-UL-us haz.loc:
Warning – Explosion Hazard – Do not disconnect while circuit is live unless
area is known to be non-hazardous.
Warning – Explosion Hazard – Substitution of components may impair
suitability for Class I, Division 2.
This equipment is suitable for use in Class I, Division 2, Groups A, B, C, D
or non-hazardous locations only.
Connection from load cells to SIWAREX FTC:
The SIWAREX FTC can be connected with load cells in zone 1 or zone 2.
For a connection of load cells in zone 2 it is sufficient, that the load cells
also have an approval for zone 2 and all associated directives as well as
the installation instructions are respected.
For a connection of load cells in zone 1, load cells with the adequate
approvals are needed and appropriate provisions according to the
guidelines must ensure explosion protected operation. For example
SIWAREX R load cells can be connected intrinsically safe via the Ex-
Interface SIWAREX IS. The instructions for the Ex-Interface, the load cells
and the ATEX as well as the c-UL-CSA directives have to be respected.
Installation and Maintenance:
Required connection and installation work must be performed by qualified
personnel.
Repairs can only be carried out at the production location.