Note
Normally, defining two operating points is sufficient to determine the characteristic
curve for the scale. Other operation points must only be defined on non-linear
systems - New digital values (adjustment digits 2, 3, 4) are assigned with other
calibration weights (e.g. 70%, 80%, 100% of the measurement range).
FIGURE 5-2
Linearization of the scale characteristic curve
The characteristic curve is converted for bi-directional operation in "Force
measurement" operating mode. The following image shows the relationship
between effective force and the adjustment digits.
FIGURE 5-3
Adjustment Digits and Force
5-38 SIWAREX FTC