Using the Pump
34
<Back OK
Pushrod must move
forward
Remove cartridge,
then press OK
Not ready
Please wait
FastForward
* In order for the pump to assure that the cartridge is correctly connected
to the pushrod, the pushrod must start in a forward position. If it is not
far enough forward when you press Load, or if you attempt to load a
cartridge that contains less than 0.2 ml, the screens shown at left will
appear. Remove the cartridge, then press OK (or press Back to return to
the previous screen). If you press OK, the pushrod moves forward, then
the screen reads, “Detach set from body, install lled cartridge, then press
Load”. If your cartridge contains less than 0.2 ml, you will need to start
with a new lled cartridge. With Load still chosen, press Select.
OK
Now pushrod must
move forward
Remove cartridge
then press OK
Sensor
conrmed
Please wait
FastForward
** If, during the loading process, you intentionally or accidentally remove
the cartridge, or if the cartridge is too full (more than 3 ml), the screens
shown at left will appear. The pump needs to check to make sure that the
cartridge sensor is working properly.
• If the cartridge is not installed, install it;
• If a cartridge is installed, make sure it is not too full (ll with no more
than 3 ml). With the tubing set attached, press down gently on the
cartridge (this will push excess medication into the tubing). Press
Conrm. The pump will conrm that the sensor is working properly. (If
the sensor is not working, you will get this message repeatedly and the
pump will need to be serviced.) If the sensor is conrmed, remove the
cartridge and press OK. The pushrod moves forward, then the screen
reads, “Detach set from body, install lled cartridge, then press Load”.
With Load still chosen, press Select.
Cancel Conrm
Cartridge missing or
overlled
Install cartridge to
conrm sensor
No cartridge
detected

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals