Command Descriptions
VX4792 User Manual
3-71
*SRE (?)
The  common command sets the bits of the SRER (Service Request Enable
Register). The  common query returns the contents of SRER.
The power-on default for the SRER is all bits reset if the power-on status flag is
TRUE. If this flag is set to FALSE, the SRER maintains its value through a
power cycle. See Status and Events for more details on the SRER.
STATUS and EVENT
, , , , ,  , !#, 
 ! "


! "

! "
where the argument must be decimal number from 0 to 255. The SRER bits are
set in binary bit according to the decimal number.
 
sets the SRER to 48 (binary 00110000), which sets the ESB and MAV bits.

might return  which indicates that the SRER contains the binary number
00100000.