Command Descriptions
VX4792 User Manual
3-77
TRIGger:OUTPut (?)
The TRIGger:OUTPut command determines whether or not the waveform
generator outputs a trigger signal onto the VXIbus. The TRIGger:OUTPut? query
returns the current setting of the trigger output.
MODE
TRIGger:INPut, CLOCk:SOURce
TRIGger:OUTPut { OFF | ECLTRIG0 | ECLTRIG1 }
TRIGger:OUTPut?
ECLTRIG1
:
OUTPut
TRIGger
<Space>
?
ECLTRIG0
OFF
OFF — no trigger is output to VXIbus
ECLTRIG0 — outputs a trigger on the ECLTRIG0 line of the VXIbus
ECLTRIG1 — outputs a trigger on the ECLTRIG1 line of the VXIbus
:TRIGGER:OUTPUT OFF
sets the trigger output off so that a trigger signal is not sent to the VXIbus. In
this example, the module becomes a trigger slave for synchronous operation.
TRIGger:POLarity (?)
The TRIGger:POLarity command selects polarity of the external trigger signal
which generates the trigger event. The TRIGger:POLarity? query returns the
currently selected polarity.
The polarity parameter is valid only when the mode is set to gated mode.
MODE
TRIGger:IMPedance, TRIGger:LEVel, TRIGger:SLOPe
Group
Related Commands
Syntax
Arguments
Examples
Group
Related Commands

Table of Contents