Introduction
This section contains a collection of tables that describe the electrical, environ-
mental, and mechanical characteristics of the VX4792 Arbitrary Waveform
Generator. The specification tables contain up to three columns: Characteristic,
Performance Requirement, and Supplemental Information. Each column is
described below.
Characteristic column. Provides the name of each characteristic.
Performance Requirement column. Describes the limits of the warranted
characteristics. Limits within this column are typically verified by performing
tests located within the Performance Verification procedure (Appendix F).
Supplemental Information column. Explains performance requirements, or
provides typical performance information.
Performance Conditions
The performance requirements found in this section are only valid under the
following conditions:
H The waveform generator must be operating as an integral part of an
appropriately configured mainframe
H The operating conditions must comply with those specified for the system
mainframe in which the waveform generator is installed
H The test environment must comply with the limits described in Table E–2
Environmental Characteristics
H The instrument must have been calibrated at an ambient temperature between
+20_ C to +30_ C following a 20 minute warm up period
H All tolerance limits apply after a 20 minute warm up period