EasyManua.ls Logo

SPECTRO XEP05 - Page 54

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance by the operator
— 54 SPECTRO XEPOS — 19.01.2017
Fault Cause Remedy
Liquid from a cuvette
has leaked into the
spectrometer.
Apparent visible
contamination.
Clean the sample chamber.
Replace the "Liquid-Protect" film of
the sample cup.
Additional contamination
that cannot be accessed.
Please contact your Service Center.
The sample chamber
lid does not open.
The measurement
procedure is not yet
complete or there is still
residual vacuum in the
sample chamber.
Wait several seconds until the end of
the measurement the sample
chamber lit is unlocked or the
sample chamber is completely
aerated.
Locking mechanism of the
sample chamber lid is
defective.
Please contact your Service Center.
The dust protection
shutter / beam
stabilization vacuum
shutter does not close
the excitation
chamber after
measurement.
The shutter is deactivated. Click the Diagnostic Tools of the
"XRF Analyzer Pro" software. Select
the menu item "Shutter" and click the
option "Re-initialize shutter".
The drive mechanism of
the shutter is defective.
Please contact your Service Center.

Table of Contents

Related product manuals