4-5
Washer Operation Operator Manual 122997-341
Once selected cycle is started, printer records:
CYCLE - CYCLE 10
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
CYCLE START 8:37:33A
CYCLE DATE 4/24/05
CYCLE NUMBER 00000001
UNIT NUMBER 3600000000
PHASE TIME F
6. Washer automatically progresses through selected cycle as fol-
lows:
NOTE: For cycle operation, note the following:
1) Time displayed on screen counts down time remaining for cycle
phase in progress. If Temperature Guarantee feature is selected,
time displayed for that phase only counts down when solution/
water temperature in chamber sump is equal to or greater than
setpoint.
2) Cycle operation may be halted at any time by pressing STOP/
RESET touch pad once. To resume cycle operation at beginning
of interrupted phase function (i.e., fill, recirculate, drain), press
CYCLE/START. To abort cycle operation, press STOP/RESET a
second time.
PRE-WASH PHASE
>> Hot water, from building supply, fills sump until required
water level is attained.
NOTE: Sump may contain rinse water retained from previous
cycle.
Display shows:
CYCLE 10 135.7f
PREWASH FILL T=2:00
>> Pre-wash water recirculates through oscillating jet system for
programmed time interval. Display shows:
CYCLE 10 139.5F
PREWASH T=1:00
Printer records time and water temperature in chamber sump at
beginning and end of recirculation:
PREWASH 10:18:15A 139.5
10:19:15A 135.0
pH NEUTRALIZATION (Option)
>> pH analyzer is initialized. Display shows:
STANDBY -
NEUTRALIZING SUMP

Table of Contents