Exhaust fan bearings must be greased once a week as follows:
1. Locate remote exhaust fan grease fittings (see Figure 6-3).
2. Apply a high temperature, water-resistant grease (P117951-
457).
NOTE: If a standard temperature grease is used, fan bearings
must be greased once at end of each day.
WARNING – ELECTRIC
SHOCK AND/OR BURN HAZ-
ARD: Disconnect all utilities
to washer before servicing.
Do not service washer unless
all utilities have been prop-
erly locked out. Always fol-
low local occupational health
and safety regulations, as
well as electric and plumbing
codes.
WARNING – BURN HAZARD:
Allow unit to cool down
before performing any ser-
vice on mechanical compo-
nents and on piping.
Components and piping
become very hot during oper-
ation.
Figure 6-3. Exhaust Fan Grease Fittings
Grease Fittings