5-8
122997-341 Operator Manual Cycle Programming
NOTE: With Temperature Guarantee feature enabled, phase
time does not count down unless setpoint is reached. It is
important that setpoint be an attainable value. Guaranteed
temperature is shown by a G before set point (GSP instead of
SP).
5. Press CHANGE VALUES touch pad. Rinse 2 phase values
screen appears:
CYCLE 10 SP=1
80.0F
RINSE 2 T= 01:30
__indicates flashing position.
6. To enter Rinse 2 temperature setpoint and phase time, press
CURSOR (left or right) touch pad to select position and VALUE
(up or down) touch pad to select desired number (0-9). Tem-
perature setpoint is input as any number value to 1/10 of a
degree. Phase time is input as minutes and seconds within a
range of 0-49 minutes and 0-99 seconds. Temperature is adjust-
able from 70 to 185°F (22 to 85°C).
7. Press CHANGE VALUES touch pad. Rinse 2 Temperature
Guarantee option screen appears:
GUAR. RINSE 2 TEMP?
NO
__indicates flashing position.
8. Press VALUE (up or down) touch pad to toggle between YES
and NO answer. Selecting NO starts Second Rinse phase time
countdown at beginning of phase. Selecting YES guarantees
that Second Rinse phase time counts down only when water
temperature is equal to or greater than programmed setpoint.
NOTE: With Temperature Guarantee feature enabled, phase
time does not count down unless setpoint is reached. It is
important that setpoint be an attainable value. Guaranteed
temperature is shown by a G before setpoint (GSP instead of
SP).
9. Press CHANGE VALUES touch pad Final Rinse phase values
screen appears:
CYCLE 10 SP=180.0F
F. RINSE T=MM
:SS
__indicates flashing position.
10. To enter Final Rinse temperature setpoint and phase time, press
CURSOR (left or right) touch pad to select position and VALUE
(up or down) touch pad to select desired number (0-9). Tem-
perature setpoint is input as any number value to 1/10 of a
degree. Phase time is input as minutes and seconds within a
range of 0-49 minutes and 0-99 seconds. Temperature is adjust-
able from 70 to 185°F (22 to 85°C).
11. Press CHANGE VALUES touch pad. Recirculated Final Rinse
screen appears:
RECIR. FINAL RINSE
NO
__indicates flashing position.
VALUE
VALUE
VALUE