6-6
122997-341 Operator Manual Routine Maintenance
6.4.3 Clean Spray Jets and
Headers
1. Observe jet nozzles during unit operation and inspect each
nozzle to detect debris.
2. Insert a wire (smaller diameter than a jet nozzle) into spray noz-
zles, and push debris through nozzles into header.
3. Remove flush plug from bottom of each spray header (see Fig-
ure 6-1).
4. Manually fill sump with water and run pump for 30 seconds to
flush loosened debris from spray headers.
5. Replace flush plugs.
Figure 6-1. Spray Header Flush Plug
Flush Plug
Spray Header