4-22
122997-341 Operator Manual Washer Operation
4.8 Bypass Phase A cycle phase may be bypassed only if Temperature Guarantee is
not selected and cycle in progress is not locked out by an access
code. Access code prevents unauthorized bypassing of phases
and/or changing cycle phase values. Refer to S
ECTION 5, PROGRAM-
MING VALUES WITH ACCESS CODE ENABLED, for details on access code
feature.
If Temperature Guarantee is not selected, cycle phase may be
bypassed only while particular phase process is in operation.
Bypassed phases apply only to immediate cycle in progress.
1. To bypass a phase, press BYPASS PHASE touch pad while
actual phase process (i.e., recirculating, exhausting) is in
operation:
CYCLE 8 181.3F
PREWASH T= 1:00
2. Control aborts phase process in operation and automatically
continues with next programmed function:
CYCLE 8 154.6F
PREWASHDRAIN T= 1:00
... and printer records:
PHASE TIME F
PRE-WASH 2:44:08P 181.3
BYPASSED PHASE 2:44:53P
DET-WASH 2:46:53P 181.4
2:51:53P 181.1
RINSE 1 2:54:48P 181.3
NOTE: Filling and draining functions can not be bypassed. If
BYPASS PHASE touch pad is pressed during these functions, control
automatically resets timer and begins filling/draining function again.
4.9 Shutdown
Procedure
At end of a work session, washer should be shut down and cleaned
thoroughly. Refer to S
ECTION 6, ROUTINE MAINTENANCE, for complete
cleaning instructions and scheduled minor maintenance.
1. Access Manual Control mode (refer to S
ECTION 4.10.1,
A
CCESSING MANUAL CONTROL MODE), and drain washer sump.
2. Position POWER-OFF/STANDBY switch to OFF/STANDBY.
3. Position building electrical disconnect switch to OFF and close
building supply valves.
4. Open manual drain valves to completely drain detergent tank
and optional acid tank, if desired.
5. Clean unit as described in S
ECTION 6, ROUTINE MAINTENANCE.
6. Ensure building electrical disconnect switch is positioned to ON
after completion of cleaning and minor maintenance proce-
dures.
NOTE: Leaving disconnect switch in OFF position overnight
shortens life span of battery backed-up control memory.
WARNING – BURN HAZARD:
When cycle is complete, par-
tially open chamber door
and allow chamber and load
to cool. Hot steam may
escape through door open-
ing if door is fully opened
after a cycle.