4-13
Washer Operation Operator Manual 122997-341
CYCLE 10 (G=) 128.5F
HEADERDRAIN T= 1.00
>> Sump water is pumped to drain. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
rinse1DRAIN T= 1.00
>> Water still present in sump is gravity-drained (option). If
temperature is guaranteed, actual sump temperature
alternates with guaranteed set temperature.
CYCLE 10 (G=) 128.5F
GRAVITYDRAIN T= 1.00
RINSE 2 PHASE
>> Hot water, from building supply, fills sump. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 137.4F
RINSEFILL T= 2.00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal to or greater than set-
point.
>> Rinse 2 solution is recirculated through oscillating jet system
for programmed amount of time. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 140.0F
RINSE2 T= 5.00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal to or greater than set
point.
Printer records time and solution temperature in chamber
sump at beginning and end of recirculation:
RINSE 2 10:21:55A 141.5
10:26:55A 140.2