4-11
Washer Operation Operator Manual 122997-341
alternates with guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
HEADERDRAIN T= 1:00
>> Sump water is pumped to drain. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
ACID DRAIN T= 1:00
>> Water still present in sump is gravity-drained (option). If
temperature is guaranteed, actual sump temperature
alternates with guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
GRAVITY DRAIN T= 1:00
>> If water saver is enabled, sump water is sent back to acid
tank (if option applies). If temperature is guaranteed, actual
sump temperature alternates with guaranteed set
temperature. Display shows:
CYCLE 10 (G=) 138.7F
ACID SAVER T= 2:00
>> If sump flush option is available, water still present in
sump is gravity-drained. If temperature is guaranteed, actual
sump temperature alternates with guaranteed set
temperature. Display shows:
CYCLE 10 (G=) 128.5F
FLUSHDRAIN T= 00.45
>> If sump flush option is available, water is injected in sump
to remove debris. If temperature is guaranteed, actual sump
temperature alternates with guaranteed set temperature.
Display shows:
CYCLE 10 (G=) 128.5F
FLUSHDRAIN T= 00.30
>> If sump flush option is available, refreshed water still
present in sump is gravity-drained. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
FLUSHDRAIN T= 00.45
RINSE 1 PHASE
>> Hot water, from building supply, fills sump. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 137.4F
RINSE1FILL T= 2.00