5-6
122997-341 Operator Manual Cycle Programming
3. Press CHANGES VALUES touch pad. Wash Temperature Guar-
antee option screen appears:
GUAR.ACID WASH TEMP?
NO
__indicates flashing position.
4. Press VALUE (up or down) touch pad to toggle between YES or
NO answer. Selecting NO starts Wash phase time countdown at
beginning of phase. Selecting YES guarantees that Wash phase
time counts down only when solution temperature is equal to or
greater than programmed setpoint.
NOTE: With Temperature Guaranteed feature enabled, phase
time does not count down unless setpoint is reached. It is
important that setpoint be an attainable value. Guaranteed
temperature is shown by a G before setpoint (GSP instead of
SP). Wash water returned to tank is shown by an R before
setpoint (RGSP or RSP).
5. Press CHANGES VALUES. Display shows Conductivity Option
Screen (if option applies):
SELECT CONCENTRATION
1/4 oz
__indicates flashing position.
6. Press VALUE (up or down) touch pad to scroll available con-
centrations (1/4, 1/2, 1, and 2 oz).
7. Press CHANGE VALUES touch pad. Display shows Time-based
Acid detergent screen, if unit is provided with a time-based acid
detergent setting:
CYCLE 10
ACID INJ. T=MM
:SS
__indicates flashing position.
8. To enter acid detergent injection time, press CURSOR (left or
right) touch pad to select position and VALUE (up or down)
touch pad to select desired number (0-9). Injection time is input
as minutes and seconds within a range of 0-99 minutes and 0-99
seconds.
9. Press CHANGE VALUES touch pad. Display shows Time-based
Acid Neutralization Option (if option applies):
CYCLE 10
ACID NEUT. T=MM
:SS
__indicates flashing position.
NOTE: If water is saved (returned to Acid tank option), question is
not displayed.
10. To enter acid neutralizer injection time, press CURSOR (left or
right) touch pad to select position and VALUE (up or down)
touch pad to select desired number (0-9). Injection time is input
as minutes and seconds within a range of 0-99 minutes and 0-
99 seconds.
VALUE
VALUE
VALUE
VALUE