122997-341 Operator Manual Troubleshooting
14. Display shows:
EMERGENCY STOP
CABLE ACTIVATED
... and printer prints:
ALARM HH:MM:SS
EMERGENCY STOP
1. Safety cable pulled – open chamber door and remove obstruction
from chamber. Press CYCLE/START touch pad to resume cycle
operation.
2. Safety cable too tight – adjust cable to allow a minimum of two
inches (50 mm) free play.
15. Display shows:
COOLDOWN TANK
TOO LONG IN FILL
... and printer prints:
ALARM HH:MM:SS
COOLDOWN TANK
TOO LONG IN FILL
Call Service.*
16. Display shows:
COOLDOWN TANK
TOO LONG TO COOL
... and printer prints:
ALARM HH:MM:SS
COOLDOWN TANK
TOO LONG TO COOL
1. Cold water supply valve not fully open – open building and
washer supply valves.
2. If situation reoccurs, call Service.*
17. Display shows:
COOLDOWN TANK
TOO LONG IN DRAIN
... and printer prints:
ALARM HH:MM:SS
COOLDOWN TANK
TOO LONG IN DRAIN
Call Service.*
18. Display shows:
COOLDOWN TANK
IS FULL
... and printer prints:
ALARM HH:MM:SS
COOLDOWN TANK
IS FULL
1. Cooldown tank high water level ball float malfunction – verify
operation of float. Clean as necessary.
2. If situation reoccurs, call Service.*
* Service Charges may be incurred. Consult your warranty for details.
Table 7-2. Operator Troubleshooting Chart - Alarm and/or Printout (Cont’d)
PROBLEM POSSIBLE CAUSE AND/OR CORRECTION