7-10
122997-341 Operator Manual Troubleshooting
24. Display shows:
pH CONTROLLER
NOT RESPONDING
... and printer prints:
ALARM HH:MM:SS
pH CONTROLLER
NOT RESPONDING
Call Service.*
25. Display shows:
CONC. TOO LOW
CHECK INJ. SYSTEM
... and printer prints:
ALARM HH:MM:SS
CONC. TOO LOW
CHECK INJ. SYSTEM
1. No alkaline detergent in container – verify and replace container.
2. No acid detergent in container – verify and replace container.
3. If situation reoccurs, call Service.*
26. Display shows:
ALK WASH
DETERGENT EMPTY
... and printer prints:
ALARM HH:MM:SS
ALK WASH
DETERGENT EMPTY
1. No alkaline detergent in container – verify and replace detergent.
2. If situation reoccurs, call Service.*
27. Display shows:
ACIDWASH
DETERGENT EMPTY
... and printer prints:
ALARM HH:MM:SS
ACIDWASH
DETERGENT EMPTY
1. No acid detergent in container – verify and replace detergent.
2. If situation reoccurs, call Service.*
28. Display shows:
CARRIAGE
SYSTEM FAILURE
... and printer prints:
ALARM HH:MM:SS
CARRIAGE
SYSTEM FAILURE
Call Service.*
* Service Charges may be incurred. Consult your warranty for details.
Table 7-2. Operator Troubleshooting Chart - Alarm and/or Printout (Cont’d)
PROBLEM POSSIBLE CAUSE AND/OR CORRECTION