Inspection Report following an Inspection of Electromedical Equipment
according to DIN EN 62353 (VDE 0751-1): 2015-10 Page 2 of 3
III.Functional test
What? How?
OK
Not
OK
Description
of defect
Functional check of the electrical components
End of travel cut-out of the motors
Automatic cut-out
Handset, control units, external power
supply
Test according to instruction
manual: Locking functions; button
function; no “rattling” when shaken
Motors
No abnormal noise development
(rattling, uneven running etc.)
Installation of cable harness and fixing
of plugs and strain relief
Securely attached, firmly fixed as
per operating instructions
Functional check of the mechanical components
Joints and pivots;
Smooth operation
Castors
Braking, secure engaging of
brakes
Safety sides
Secure engagement, unlocking acc.
to operating instructions
Lower leg rest
Engages properly
Accessories (e.g. patient lifting pole,
grab handle)
Secure attachment, without
damage, suitability for bed
Inspection result:
Inspection passed; test approval sticker applied:
No safety-related or functional defects were detected
No direct risk; the defects detected can be rectified quickly
Inspection was not passed; no test approval sticker applied:
Appliance must be taken out of circulation until the defects have been rectified!
Device does not meet requirements modification / replacement of components / decommissioning is
recommended!
Next inspection date:
Documents that form part of this inspection report:
Enclosure, page 3/3: Dimensional check of safety sides in compliance with statutory regulations
Inspected on:
Inspected by:
Signature:
Evaluated on:
Operator/Expert:
Signature: