• Do not leave a patient unattended. Hold the product while a patient is on the product.
• Do not apply a wheel lock when a patient is on the product or when you move the product to avoid the risk of a cot tip or
injury.
• Always secure the mattress to the product to avoid movement during patient transfer.
CCAAUUTTIIOONN -- Do not use siderails as a patient restraint device.
To transfer the patient to the cot:
1. Roll the cot to the patient (
Rolling the cot with a patient
(page 30)).
2. Place the cot beside the patient. Raise or lower the cot to the level of the patient.
3. Lower the siderails and open the restraint straps.
4. Transfer the patient to the cot. Follow accepted EMS procedures.
5. Secure the patient to the cot with all of the restraint straps.
6. Raise the siderails and adjust the backrest and footrest as necessary.
RRaaiissiinngg oorr lloowweerriinngg tthhee ccoott
You can raise or lower an unoccupied cot with one operator. If a patient is on the cot, a minimum of two trained operators
are required to raise or lower the cot (one located at each end of the cot).
WWAARRNNIINNGG
• Always keep hands, fingers, and feet away from moving parts. Use caution when you place your hands and feet near the
base tubes to raise or lower the cot.
• Always follow proper hand placement on hand grips. Keep all hands clear of the red safety bar pivots when you load,
unload, or change the cot height position.
• Do not raise or lower the cot while you are on rough or uneven terrain.
There are two identical cot control switches located on your PPoowweerr--PPRROO 2 cot. Press the button on either of these switches
to raise (extend) the cot, lower (retract) the cot, or release the cot from PPoowweerr--LLOOAADD, if applicable (Figure 10).
FFiigguurree 1100 –– CCoott ccoonnttrrooll sswwiittcchheess
RReeffeerreennccee NNaammee DDeessccrriippttiioonn
A Foot end light Press to turn the foot end light on or off
B Indicator light Illuminates when you reach transport
height (see item D)
C Retract (-) Press and hold to lower the litter or
retract the cot undercarriage
EN 28 6507-009-001 Rev AB.3