CClleeaanniinngg tthhee pprroodduucctt
WWAARRNNIINNGG
• Always follow these cleaning and disinfecting guidelines, in addition to your protocols, to maintain hygienic safety.
• Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power
washing equipment may aerate contamination.
CCAAUUTTIIOONN
• Always remove the battery before you wash the product.
• Do not clean, service, or perform maintenance while the product is in use.
• Do not steam clean or ultrasonically clean the product.
• Do not exceed 180 °F (82 °C) as the maximum water temperature.
• Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you are using a hand held wand to wash the
product, keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
The product is power washable. The product may show some signs of oxidation or discoloration from continuous washing.
No degradation of the product’s performance will occur from power washing as long as you follow the proper procedures.
• Immersing restraint strap metal buckles can cause buckle corrosion and is not recommended. Rinse with clean water
and allow to air dry to reduce chance of corrosion. Replace restraints if metal buckles are corroded.
• Direct skin contact with visibly soiled, permeable material may increase the risk of infection.
Recommended cleaning method:
1. Remove the mattress and battery from the product.
2. Follow the disinfectant solution manufacturer’s dilution recommendations exactly.
3. Stryker recommends the standard hospital cart washer for power washing.
4. Allow the product to air dry.
5. Check functionality before you return the product to service.
6507-009-001 Rev AB.3 57 EN