EEMMCC iinnffoorrmmaattiioonn
WWAARRNNIINNGG
• Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be used no closer than 12 inches (30 cm) to any part of PPoowweerr--PPRROO 2, including cables specified by the manufacturer.
• Avoid stacking or placing other equipment adjacent to PPoowweerr--PPRROO 2 to prevent improper operation of the products. If
such use is necessary, carefully observe the cot and the other equipment to verify proper operation.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc eemmiissssiioonnss
EEmmiissssiioonnss tteesstt
CCoommpplliiaannccee EElleeccttrroommaaggnneettiicc eennvviirroonnmmeenntt
PPoowweerr--PPRROO 2 is intended for use in the electromagnetic environment specified below. The customer or the user of
PPoowweerr--PPRROO 2 should assure that they are used in such an environment.
RF emissions
CISPR 11
Group 2
PPoowweerr--PPRROO 2 with the PPoowweerr--LLOOAADD
compatibility option must emit electromagnetic
energy in order to perform its intended
function. Nearby electronic equipment may be
affected.
RF emissions
CISPR 11
Group 1
The emissions characteristics of this
equipment make it suitable for use in
professional healthcare facilities, emergency
medical services, and home healthcare
environments. If it is used in other
environments, this equipment might not offer
adequate protection to radio-frequency
communication services and power supply
networks. The user might need to take
mitigation measures, such as relocating or
reorienting the equipment.
RF emissions
CISPR 11
Class B
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
PPoowweerr--PPRROO 2 is suitable for use in a professional healthcare facility, home, and EMS environments. PPoowweerr--PPRROO 2 is
not suitable for use in environments exceeding immunity test conditions that the product was evaluated to, such as near
high frequency (HF) surgical equipment and inside of the radio frequency (RF) shielded room of magnetic resonance
imaging (MRI) equipment. The customer or the user of PPoowweerr--PPRROO 2 should assure that it is used in such an
environment and that the electromagnetic environment guidance listed below is followed.
IImmmmuunniittyy tteesstt
IIEECC 6600660011 tteesstt lleevveell
CCoommpplliiaannccee lleevveell
EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt--gguuiiddaannccee
Electrostatic discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
6507-009-001 Rev AB.3 65 EN