EasyManua.ls Logo

Stryker Power-PRO 2 - Every six months or 12 hours; Every 12 months or 24 hours

Stryker Power-PRO 2
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EEvveerryy ssiixx mmoonntthhss oorr 1122 hhoouurrss
Inspect these items every six months or 12 hours, whichever comes first.
IItteemm IInnssppeecctt
Electronic controls/functions Extend cot to raised position, measure and check load
height
Jog function operates
High speed retract and extend operates
Bumper detection operates
Press the release or transport height button and confirm
correct height
Measure load height and confirm correct height
Switches No damage or wear to the switches
All switches operate
Litter No bent, broken, or damaged components
No damage or tears on cot grips
Siderails operate and latch
Footrest operates
Mattress No cracks or tears
Base No bent, broken, or damaged components
Cot retaining post is secure. If not, replace the screw.
No excessive damage to X-frame guards
Wheels Free of debris
SStteeeerr--LLoocckk and wheel locks operate
Check brake cable (between SStteeeerr--LLoocckk and wheel lock)
for wear, bends, creases
Head section No bent, broken, or damaged components
Grip bar has no excessive damage or tears
Load wheels are secure and roll
Foot section No bent, broken, or damaged components
Grip bar has no excessive damage or tears
EEvveerryy 1122 mmoonntthhss oorr 2244 hhoouurrss
Inspect these items every 12 months or 24 hours, whichever comes first.
IItteemm IInnssppeecctt
Settings Cot and fastener fit and function
Safety bar connects to the vehicle safety hook
EN 62 6507-009-001 Rev AB.3