DDiissiinnffeeccttiinngg tthhee pprroodduucctt
In general, when used in concentrations recommended by the manufacturer, either phenolic type or quaternary (excluding
VViirreexx® TB) type disinfectants can be used. Iodophor type disinfectants are not recommended for use because staining
may occur.
Recommended disinfectants:
Quaternary cleaners (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic cleaners (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
21% isopropanol alcohol
Recommended disinfection method:
1. Follow the disinfectant solution manufacturer’s dilution recommendations exactly.
2. Hand wash all surfaces of the product with a disinfectant solution.
3. Avoid oversaturation and make sure that the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting.
4. Allow the product to air dry.
5. Disinfect the VVeellccrroo® after every use. Saturate the VVeellccrroo® with disinfectant, rinse with water, and allow the
disinfectant to evaporate. Appropriate disinfectant for nylon VVeellccrroo® should be determined by the service.
6. Check functionality before you return the product to service.
NNoottee
Failure to follow the above directions when using these types of disinfectants may void this product’s warranty.
Always wipe the product with clean water and dry after cleaning. Some cleaning products are corrosive in nature and
may cause damage to the product. Failure to properly rinse and dry the product leaves a corrosive residue on the
surface of the product and may cause premature corrosion of critical components.
EN 58 6507-009-001 Rev AB.3