CyFlow™ Cube 6 V2m | Operating Manual | February 2020
Cleaning Cycles
The different measure modes handle the cleaning differently
• [Analyze all] and [VC Electrodes] include an automatic cleaning cycle after
sample acquisition. The cleaning cycle will not start, if the measurement is
terminated with the [End] button.
• [VC of Volume] and [Cells in Region] do NOT include an automatic
cleaning cycle after sample acquisition. Intermediate cleaning is required.
Empty sample
The sample analysis automatically stops when the sample is empty, even
if the selected end criterion is not yet reached.
Measure Mode allows the selection of different start and end criteria for the measurement.
Figure 33:Acquisition process for all Sample Port Measure Modes
Sample Volume
Always use additionally to the sample material a dead volume of 50 µl.
VC of Volume
Cells in Region