11 Software
62
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
Table 17: Manual Sample Port Measure Modes
Measure Mode
Description
VC Electrodes
True Volumetric Absolute Counting. The start and stop
electrode are used to measure a defined volume of 200 µl.
VC of Volume
An adjustable predefined volume is analysed.
Analyze all
Continuous acquisition of the complete sample.
Cells in Region
Continous acquisition until a flexible predefined number of
particles within in a specified region is reached.
Parameters of Workstep
Speed and Volume values can be stored by saving a configuration file.
Figure 34: Workstep register
Property
Options
Description
Command +
Parameter
Flow
Adjusts the sample flow [µl/s] for pre-run and
measure phases
Volume
Adjusts the measured volume [µl] for pre-run and
measure phases