5. Error Messages and Trouble Shooting
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 5-3
Error Messages Description Trouble Shooting
Filter Errors
Already inserted Filter already inserted Check if the filter is properly inserted.
No filter carriage detected The instrument does not
detect the filter carriage
Insert filter. If a filter has been already
inserted, check the filter carriage for
dirt or damage. Contact your local
service engineer, if the error continues.
No measurement filter
defined
The measurement filter is
not defined
Define filter.
No reference filter defined The reference filter is not
defined
Define filter.
Illegal filter carriage
position
Internal firmware or
electrical error
Check the filter carriage for dirt or
damage. Contact your local service
engineer, if the error continues.
Wavelength … nm not
available
The defined reference or
measurement filter is not
available on the inserted
filter carriage
Change filter carriage or check filter
values for incorrect input.
Filter carriage not defined,
Type.. Number…
Wrong, damaged or not
defined filter carriage
inserted
Check the filter carriage to see if it is
correct or check the filter carriage for
dirt and damage.
ADC Electronic Error
Offset 340 not adjusted Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.
Offset 400 not adjusted Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.
No wavelength defined Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.
Area 400 not adjusted Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.
Area 340 not adjusted Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.
E2Pot Overflow Electronic error on ADC
board or optical problem
Start Lamp Adjust program in the
Setup program. Contact your local
service engineer, if the error continues.