5. Error Messages and Trouble Shooting
TECAN
5-4 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
5.1.2 Table of Error Messages and Trouble Shooting for
SPECTRA Mode
Error Messages Description Trouble Shooting
Filter The USB stick has not
reached the required
position or is not defined.
Check that the USB stick is correctly inserted
into the instrument. Contact your local
service engineer, if the error continues.
Transport Microplate transport error Ensure that the microplate is inserted
correctly and that nothing is blocking the
transport system. Contact your local service
engineer, if the error continues.
Lamp low Optical system error This message is displayed when the
instrument detects that the optical system is
not receiving enough light. Possible causes
could be: defective halogen lamp, halogen
lamp incorrectly positioned, optical system is
not clean, filter alignment out of range.
Contact your local service engineer, if the
error continues.
Lamp high Optical system error This message is displayed when the
instrument detects that the optical system is
receiving too much light. Possible causes
could be: defective halogen lamp, halogen
lamp incorrectly positioned. Contact your
local service engineer, if the error continues.
System Internal firmware or flash
EPROM error
Contact your local service engineer.
Abort Lamp or other optical
defect
Check lamp and if the lamp is working and
positioned correctly, contact your local
service engineer.