6. Maintenance & Cleaning
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 6-11
Caution
Make sure that the microplate has been removed from the instrument
before starting disinfection.
Observe the following issues during disinfection:
1. Wear protective gloves, protective glasses and protective clothing.
2. Prepare a suitable container for all disposables used during the disinfection
procedure.
3. Move the plate transport into the load position.
4. If applicable, remove the microplate from the plate transport.
5. Disconnect the instrument from the main power supply and let it cool down
to ambient temperature to avoid any risk of fire and explosion.
6. Disconnect the instrument from the computer.
7. Carefully apply the disinfectant solution according to the manufacturers
instructions for use on the plate transport of the instrument.
8. After the required contact time (according to the manufacturers instructions
for use) wipe the plate transport using a soft paper towel moistened with a
mild detergent or distilled water to remove all traces of the disinfectant.
9. Move the plate transport into the instrument by gently pressing the front end
of the plate transport until the plate transport door is completely closed.
10. Carefully apply the disinfectant solution according to the manufacturers
instructions for use on all outer surfaces of the instrument.
11. After the required contact time (according to the manufacturers instructions
for use) wipe the instrument using a soft paper towel moistened with a mild
detergent or distilled water to remove all traces of the disinfectant.
12. Wipe dry the outer surface of the instrument with a soft paper towel.
13. Disinfect your hands and clean them with a mild detergent.
14. Pack the instrument and its accessories.
15. Dispose of the container with the disposables according to the relevant
national laws and regulations.
16. Complete the disinfection declaration and attach it to the outside of the box
so that it is clearly visible.
Caution
The plate transport should only be moved manually if the instrument is
disconnected from the main power supply.
See below for the disinfection declaration, which must be completed before the
instrument is returned to the distributor/ service center.