INSTRUMENT DESCRIPTION
Table 1.2 summarizes the accuracy specifications for the model 1750, in the delayed mode.
RANGE REFERENCE CURRENT
1 A 100 mA 10 mA 1 mA 100 μA 10 μA 1 μA 100 nA
2
.02%+5ct
20
.02%+4ct
.02%+5ct
200
.02%+2ct
.02%+4ct
2 Ω
.02%+2ct
.02%+4ct
20 Ω
.02%+2ct
200 Ω
.02%+2ct
.02%+4ct
2
.02%+2ct
20
.02%+2ct
.02%+2ct
200
.02%+2ct
2
.04%+2ct
20
.04%+2ct
Table 1.2: Delayed Mode Accuracies with Respect to Reference Current
(±) ACCURACY; 18-28 °C (64.4-82.4 °F); 1 YEAR (after initial 6 month cycle).
In the fast mode, the published accuracy for all reference current ranges is +/- (0.05% + 5 counts).
These accuracies are displayed in Table 1.3. Fast Mode is not available for all ranges. If the unit is in
Fast Mode and a range is selected that Fast Mode is not available, then the instrument will default to
Delayed mode.
RANGE REFERENCE CURRENT
1 A 100 mA 10 mA 1 mA 100 μA 10 μA 1 μA 100 nA
2
NA
20
NA
NA
200
.05%+5ct
NA
2 Ω
.05%+5ct
NA
20 Ω
.05%+5ct
NA
200 Ω
.05%+5ct
.05%+5ct
NA
2
.05%+5ct
.05%+5ct
20
.05%+5ct
NA
200
NA
2
NA
20
NA
Table 1.3: Fast Mode Accuracy with Respect to Reference Current