Power Up and Device Clear Defaults
When the 1750 is first powered up the machine state reinitializes to the last settings it had before its
last power down. When a device clear or selected device clear is executed from the bus, the machine
state reinitializes to factory defaults. See Table 3-1 for a summary of these default settings.
Reading from the IEEE-488 bus
Below are some BASIC instructions for making a 1 measurement of a resistor via the interface bus.
This will be executed by programming the 1750 to the 2 kΩ @ 1 mA range, fast one-shot trigger, and
take a resistance reading.
Send the ASCII string, “R13T1X” to address 12
[OUTPUT 12; “R13T1X”
Where R13 = 2 @1 mA range
T1 = Set 1750 to Fast One-Shot Trigger
X = Message Delimiter
Retrieve the reading of ASCII string “1.0000 kOhm” with default terminators of <CR><LF>and EOI
ENTER 12
LINE INPUT A$
PRINT A$