MISCELLANEOUS SPECIFICATIONS (Continued)
Dimensions:
Depth: 13.0” (33.0 cm)
Width: 8.50” (21.6 cm)
Height: 5.20” (13.2 cm)
Weight:
9.25 lbs (4.20 kg)
Rear Interface Relay Contact Specifications:
The relay contact I/Os are rated at 125 VAC @ 500 mA or 30 VDC @ 1 A.
Calibration:
Calibration of the 1750 is permitted via the front panel with no internal adjustments.
Calibration requires the temporary placement of a jumper, P9 to the J9 position.
AC Power Requirements:
Input: <100 VA, 108-132 VAC or 216-264 VAC, at 50/60 Hz.
Fuse:
For 108-132 V Operation; use 0.8 A @ 250 V, 5X20 mm, fast acting, TEGAM PN#FU-800
For 216-264 V Operation; use 0.5 A @ 250 V, 5X20 mm, fast acting, TEGAM PN#FU-500