EasyManua.ls Logo

Tegam 1750 - Page 7

Tegam 1750
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-3
INSTRUMENT DESCRIPTION
Auto- Correct Function
The auto- correct function compensates for time or temperature drift by reading high accuracy
internal references and automatically calculating correction coefficients for all ranges. During this
software correction process, the 1750 will delay one reading for 50 ms. The auto correct routine
is performed at power up, after 30 seconds of operation, after 30 minutes of operation, and after
every 65,536 “delayed trigger” readings, (approximately 2 hours).
High Noise Immunity
Built in line cycle integration in combination with shielded test leads produce exceptional noise
rejection.
GPIB, RS232, or RS-422 Communication Interfaces
Standard GPIB, RS232, or RS422 user-selectable interfaces are standard communications for all
1750 units. A simple command sets ease the task of integration while the use of device dependent
commands allows for specialized function execution.

Table of Contents

Related product manuals