Operating Instructions
1502C MTDR Service Manual
1–27
8. To exit monitor mode, push STORE again.
9. To exit Max Hold, access the Acquisition Control Menu again, turn off Max
Hold, and push MENU repeatedly until the instrument returns to normal
operation.
This feature puts the 1502C in a “listening mode” by turning off the pulse generator.
1. Attach a cable to the 1502C front-panel CABLE connector.
2. Push MENU to access the Main Menu.
3. Scroll to Setup Menu and push MENU again.
4. Scroll to Acquisition Control Menu and push MENU again.
5. Scroll to Pulse is: On and push MENU again. This will change to Pulse is: Off.
O
F
F
O
F
F
O
F
F
O
N
ac 0.000 ft
Figure 1–32: Waveform Display with No Outgoing Pulses
6. Repeatedly press MENU until the instrument returns to normal operation.
CAUTION. This function is used mostly for troubleshooting by qualified technicians.
It is not recommended that you use the 1502C as a stand-alone monitoring device.
The input circuitry is very sensitive and can be easily damaged by even moderate
level signals.
NOTE. In this mode, the 1502C is acting as a detector only. Any pulses detected will
not originate from the instrument, so any distance readings will be invalid. If you
are listening to a local area network, for example, it is possible to detect traffic, but
not possible to measure the distance to its origin.
Pulse is: Off can be used in conjunction with Max Hold is: On.
7. To exit Pulse is: Off, access the Acquisition Control Menu again, turn pulse back
on, then push MENU until the instrument returns to normal operation.
Pulse On/Off