EasyManuals Logo

Tektronix TG8000 User Manual

Tektronix TG8000
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
Operating basics
RECALL. Recall
s instrument settings saved in the selected preset. Use the left
()orright() arrow button to select a different preset, and then press the
ENTER button to implement the selection. While the preset is loading into the
instrument, the name of the selected p reset blinks. When the loading is complete,
the blinking stops.
SAVE. Saves
the current instrument settings to one of the 14 available presets.
Use the left ()orright() arrow button t o select the destination preset, and
then press the ENTER button to implement the selection.
RENAME. Re names the selected preset. Perform the follow i ng steps to rename
a preset:
1. Use the left ()orright() arrow button to select the preset you want to
rename.
2. Press the ENTER button twice to begin editing the preset name. The
underscore character (_) appears under the rst letter of the preset name.
3. Use the left ()orright() arrow button to move the underscore character
to the letter you want to modify.
4. Use the up ()ordown() arrow button to select the new character. The
available characters include the full alphabet, numerals from 0 to 9, and many
stan
dard ASCII symbols. The text strings may contain up to 32 characters.
5. Enter all of the desired characters then press ENTER button.
6. Press the BACK button twice to exit the PRESET submenu. This returns to
the top of the m ainframe main menu.
7. Press the ENTER button to reenter the PRESET submenu.
8. Use the up ()ordown() arrow button to select the SAVE menu item.
9. Use the left () or right () arrow button to select the preset name edited
in step 2.
10. Press ENTER button to save the preset with the new name.
DELETE. Deletes the contents of the specied preset. Use the left ()orright
() arrow button to select the preset you wa nt to delete, and then pres s the
ENTER button to implement the selection. You cannot d elete the following
presets: Power on Default, User Default, and Factory Default.
2–24 TG8000 Multiformat Test Signal Generator User Manual

Table of Contents

Other manuals for Tektronix TG8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TG8000 and is the answer not in the manual?

Tektronix TG8000 Specifications

General IconGeneral
BrandTektronix
ModelTG8000
CategoryPortable Generator
LanguageEnglish

Related product manuals