Versa User Manual www.teledynetekmar.com
Introduction • Page 1-7
Therefore, the three variables, temperature, pressure and volume, have a direct aect on the sensitivity of the system.
The temperature in the Versa that would aect injection mass is the Valve Oven Temperature. The lower the temperature
used, the denser the mass of the gas in the loop. The pressure aects the sensitivity in two areas: the higher the vial
pressurization setting, the more dilution gas is added thus lowering sensitivity; while a higher loop ll pressure results in
higher sensitivity. The volume of the loop can be changed to allow for more mass to be injected on column. The volume
can be varied between the optional loop sizes of 100μl up to 5mL. The Versa is shipped with a 1mL loop.
Note: Refer to the Method Development section of this manual for further information on the eects of temperature,
pressure and volume.
Injection
Once the Sample Loop has been lled, and the GC is ready for injection and the 6-port valve is rotated. When the 6-port
valve is in the “B” position, it allows carrier gas to back ush the sample from the loop onto the analytical column. The GC
data system is started concurrently. The vial is lowered from the Platen and the sample pathway is subsequently swept
while waiting on the equilibration of the next sample to eliminate cross contamination.
SYSTEM SPECIFICATIONS
Technical Specications
Autosampler Capacity 20 Positions
Vial Size 22 mL vials
Platen Heater Single position; ambient to 200° C via resistance heater settable in 1° C increments
Sample Loop Ships with a standard 1mL Silco® coated loop. Other optional loop sizes include 100, 250, 500µL as
well as 2, 3, and 5mL
Sample Mixing System allows variable power settings from low/medium/high
Sample Pathway Siltek® tubing. Entire sample pathway temperature controlled up to 200° C
Software TekLink™2G software in a Windows® XP or greater environment via USB connnection. TekLink™2G can
be congured for full 21 CFR Part 11 compliance tool capability
GC Interface Interfaces to virtually all commercially available GC instruments. Supplies or accepts GC and Data
System start/ready signals via software selectable GC I/O conguration
Pressure Control Software has the ability to control pressure to assure consistent volume regardless of external
conditions for all samples via a pressure transducer
Valving Motor actuated 6-port valve with removable rotor. Temperature controlled from ambient to 200° C.
Voltage Requirements 100/115 VAC +/- 5%, 50/60 Hz, 5 amps & 220/240 VAC +/- 5%, 50/60 Hz, 2.5 amps
Gas Requirements Ultra-high purity (99.999%) Helium or Nitrogen; incoming pressure: 45-140 psi
Unit Dimensions 21” (53.34cm) H x 12” (30.48cm) W x 21” (53.34cm) D
Weight 40lbs (18.14kg)
Environmental
Conditions
Operating Temperature: 10° to 30°C; Storage Temperature; -20° to 60° C;
Relative Humidity: 10% to 90%
Corrosion Resistance The chassis and carousel tray are resistant to aqueous samples with a pH range of 1-10
Certications CE, CETL, CSA, ETL
Service
Electronic Leak Check Ability to leak check the sample pathway of the system via an automated system leak check process.
Benchmark Test The system has a mode that will allow for full electromechanical testing including; valving, heater, vial
handling systems, liquid delivery system, inputs and outputs.
Diagnostics
The system oers independent control of all valves and vial handling mechanisms and for troubleshooting.
E-mail Alert The system can be congured to send an E-mail to alert the user of schedule completion or stoppage.
Warranty 1 year from installation. 18 months from shipment.