Teledyne Tekmar Versa - Prepare Vials for Sampling; Spiking a Vial

Teledyne Tekmar Versa
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Versa User Manual www.teledynetekmar.com
Installation & Setup • Page 3-4
Prepare Vials for Sampling
1. Place the cap on a rm surface.
2. Insert the septum into the cap (avoid contamination of the septum face, the shiny Teon® face should always be
face-down toward the vial).
3. Place the septum and cap together on top of the vial opening.
4. Place the crimper on top of the capped vial with the crimp head resting on the cap.
5. Squeeze the crimper handles rmly in a slow and smooth motion, do not squeeze quickly.
6. Remove the crimper.
7. Check for a proper crimp seal by lightly rotating the cap. If some resistance exists, it is properly sealed.
Spiking a Vial
Spiking a vial containing a fortication solution such as an internal standard, surrogate standard, or stock standard can
be done prior to capping or after the vial is sealed. Both techniques have advantages and disadvantages.
Figure 3-4: Hand Crimper and Vial
Septum
(Teon® face
down toward
vial opening)
Cap
Sample Vial
Crimper