Remote Interfaces
TP04100A Interface
&
Applications
Rev A: 06/1
5/98
4-8.1 Mandatory
IEEE
488.2 Common Commands and
Queries
*CLS Clear the status (*ESR, TESR) registers
i
a
*ESE
Set the standard event status enable (mask) register
*ESE nnn
--
where nnn is 0
-
255
*ESE?
Read the standard event status enable (mask) register
I
*ESR? Read the standard event status register
bit
7
--
power on
--
NOT USED
6
--
user request
--
NOT USED
5
--
command error (cme)
4
--
execution error (exe)
3
--
device dependent error (dde)
2
--
query error (qye)
1
--
request control
--
NOT USED
0
--
operation complete
--
NOT IMPLEMENTED
NOTE:
The above bits are latched and are automatically cleared when the register is read.
*GL Go to local
*IDN? Read the product identification information (Manufacturer, model, software version). Returns
TEMPTRONIC, TP04100A, PG129500A. The software version field in the return string will
vary with the particular software version.
*LL Local Lockout
*RST
Reset the device-specific functions ("Program" mode). The TP04100A system is forced to
the Manual Operation Screen, with setpoint number 1 active. Any device-specific errors are
reset. If the system
was
in program mode, the software configuration file, and the currently
chosen test setup (including setpoints) are loaded from memory. If the system was already
in
manual mode, the current configuration, setup and setpoint information are left unchanged.
*SRE
Set the service request enable (mask) register
*SRE
nnn
--
where
nnn
is 0
-
255.
*SRE? Read the service request enable (mask) register
Page 4-1 0