9 Operation
82
Click on the Measurement parameters tab.
Click on the required measurement parameter.
Store or edit the relevant data.
All changes are automatically stored in the respective probe.
9.7.3 Surface increment
Surface probes withdraw heat from the measured surface immediately
after the initial contact. This makes the measurement result lower than
the true surface temperature without the probe (or the reverse if the
surface is colder than the environment). This effect can be corrected by
an increment in % of the reading.
The probe is connected to the testo 400.
Click on .
Main menu opens.
Click on Sensors.
The Sensors menu opens.
The TC type K probes are not displayed as separate probes, but always
on the instrument itself.
Information is displayed about the model, order number, serial number
and firmware version.
Click on the Measurement parameters tab.
Window with measurement parameters opens.
Click on the required Temperature TE1 or Temperature TE2.
Window for surface increment opens.
Click on Activate surface increment.