9 Operation
86
9.7.6 Humidity calibration
For humidity calibration, the measurement parameter of the connected
probe is adjusted to the reference value at the two
points 11.3 %RH and 75.3 %RH; any measuring value deviations from
the nominal value are minimized across the entire measuring range.
The Testo calibration set provides a reference value to calculate the
offset for a humidity calibration.
Humidity calibration is possible with the following probes:
Order no. Designation
High-precision temperature/humidity probe with Bluetooth
®
High-precision temperature-humidity probe, fixed cable
Temperature/humidity probe with Bluetooth
®
Temperature/humidity probe, fixed cable
Robust temperature/humidity probe for temperatures up to
+180 °C, fixed cable
The probe is connected to the testo 400.
Click on .
Main menu opens.
Click on Sensors.
The Sensors menu opens.
Click on the required sensor.
Information is displayed about the model, order number, serial number
and firmware version.
Click on the Measurement parameters tab.
Window with measurement parameters opens.
Click on Relative humidity.
The Relative humidity window opens.